KL25-0.6/1.5-110 stationary anode x-ray tube is especially designed for C-arm equipment and available for nominal tube voltage with high frequency or DC generator.
The integrated high quality tube with glass design has two super imposed focal spot and a reinforced anode. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.
KL25-0.6/1.5-110 is especially designed for C-arm equipment and available for nominal tube voltage with DC generator.
This tube has focus 1.5 and 0.6 foci and is available for maximum tube voltage 110 kV.
| Nominal Tube Voltage | 110kV |
| Nominal Focal Spot | small:0.6 large:1.5(IEC60336/2005) |
| Filament Characteristics | small:Ifmax=4.5A,Uf=5±0.5 large:Ifmax=4.5A,Uf=6.3±0.8V |
| Nominal Input Power(at 1.0s) | small:spot0.6kW large:spot3.5kW |
| Maximum Continuous Rating | 225W |
| Anode Heat Storage Capacity | 30kJ |
| Target Angle | 12° |
| Target Material | Tungsten |
| Inherent Filtration | Min 0.6mmAl equivalent at 75kV |
| Weight | approx540g |

Environmental Limits
Operating Limits (in dielectric oil):
Oil Temperature ……………………………………………………………………………………10 ~ 60 °C
Oil Pressure ……………………………………………………………………………………. 70 ~ 106 kPa
Shipping and Storage Limits: Temperature………………………………………….-40~ 70 °C
Humidity …………………………………………………………………………………………….. 10 ~ 90 %
(No condensation)
Atmospheric Pressure ……………………………………………………………………… 50 ~ 106 kPa
Elevated anode heat storage capacity and cooling
Constant high dose yield
Excellent lifetime
Minimum Order Quantity: 1pc
Price: Negotiation
Packaging Details: 100pcs per carton or customized according to the quantity
Delivery Time: 1~2 weeks according to the quantity
Payment Terms: 100% T/T in advance or WESTERN UNION