This product has been manufactured and developed in agreement with the following laws,directives and design regulations:
◆Council Directive 93/42/EEC of 14 June 1993 concerning medical devices(CE marking).
◆EN ISO 13485:2016 Medical device—Quality management systems—Requirements for regulatory
purposes..
◆EN ISO 14971:2012Medical devices – Application of risk management to medical devices (ISO 14971:2007, Corrected version 2007-10-01)
◆EN ISO15223-1:2012Medical devices——Symbols to used with medical device labels, labeling and information to be supplied Part 1: General requirements
◆International Electrotechnical Commission (IEC),the following standards are considered in particular.
Standard Reference |
Titles |
EN 60601-2-54:2009 | Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy |
IEC60526 | High-voltage cable plug and socket connections for medical X-ray equipment |
IEC 60522:1999 | Determination of the permanent filtration of X-ray tube assemblies |
IEC 60613-2010 | Electrical, thermal and loading characteristics of rotating anode X-ray tubes for medical diagnosis |
IEC60601-1:2006 | Medical electrical equipment – Part 1: General requirements for basic safety and essential performance |
IEC 60601-1-3:2008 | Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment |
IEC60601-2-28:2010 | Medical electrical equipment – Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis |
IEC 60336-2005 | Medical electrical equipment-X-ray tube assemblies for medical diagnosis-Characteristics of focal spots |
●The designation is composed as follows:
MWHX7010 |
Tube |
A |
High voltage socket with 90 degree direction |
MWTX70-1.0/2.0-125 |
B |
High voltage socket with 270 degree direction |
Property |
Specification |
Standard |
|
Nominal input power(s) of the anode |
F 1 |
F 2 |
IEC 60613 |
21kW(50/60Hz) |
42.5kW(50/60Hz) |
||
Anode heat storage capacity |
100 kJ ( 140kHU) |
IEC 60613 |
|
Maximum cooling capacity of the anode |
475W |
||
Heat storage capacity |
900kJ |
||
Max. continuous heat dissipation without Air-circular |
180W |
||
Anode materialAnode top coating material |
Rhenium-Tungsten-TZM(RTM) Rhenium-Tungsten-(RT) |
||
Target angle (Ref: reference axis) |
16 ° |
IEC 60788 |
|
X-ray tube assembly inherent filtration |
1.5 mm Al / 75kV |
IEC 60601-1-3 |
|
Focal spot nominal value(s) |
F1(small focus) |
F2(large focus) |
IEC 60336 |
1.0 |
2.0 |
||
X-ray tube nominal voltageRadiographic
Fluoroscopic |
125kV 100kV |
IEC 60613 |
|
Data on cathode heating
Max. current Max voltage |
≈ /AC, < 20 kHz |
||
F1 |
F 2 |
||
5.1A ≈5.8~7.8V |
5.1 A ≈7.7~10.4 V |
||
Leakage radiation at 150 kV / 3mA in 1m distance |
≤ 1.0mGy/h |
IEC60601-1-3 |
|
Maximum radiation field |
573×573mm at SID 1m |
||
X-ray tube assembly weight |
Approx. 18 kg |
Limits |
Operation Limits |
Transport and Storage Limits |
Ambient temperature |
From 10℃to 40℃ |
From- 20℃to 70℃ |
Relative humidity |
≤75% |
≤93% |
Barometric pressure |
From 70kPa to 106kPa |
From 70kPa to 106kPa |
1-phase stator
Test point |
C-M |
C-A |
Winding resistance |
≈18.0…22.0Ω |
≈45.0…55.0Ω |
Max.permissible operating voltage(run-up) |
230V±10% |
|
Recommend operating voltage(run-up) |
160V±10% |
|
Braking voltage |
70VDC |
|
Run-on voltage in exposure |
80Vrms |
|
Run-on voltage in fluoroscopy |
20V-40Vrms |
|
Run-up time(depending on starter system) |
1.2s |
Warning to interface with X-ray Generator
1.Housing Rupture
Never input over rated power to X-ray tube assembly
If the input power exceeds tube specification ,it causes the over temperature of anode,insert tube glass shatter and ultimately the following serious problems due to generating over-pressure by oil vaporization inside housing assembly.
In such a critical condition causing housing rupture by over load ,the safety thermal switch can not protect X-ray tube even if it work
*Housing sealing parts rupture.
*Human injury including burns due to hot oil escape.
*Fire accident due to flaming anode target.
The X- ray generator should have a protective function which manages input power to be within tube specification.
2.Electric Shock
To avoid the risk of electric shock ,this equipment must only be connected to a supply with protective earth.
3.No modification of this equipment is allowed!!
Caution to interface with X-ray Generator
1.Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical date sheets carefully and follow the instructions.
2.Permanent Filtration
The total filtration and the distance between X-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
3.Safety Thermal Switch
X-ray tube assembly has safety thermal switch to prohibit further input power when the tube housing reaches to the temperature(80℃) of switch-open.
The switch is not recommended connecting stator coil in series circuit.
Even if the switch works ,never turn the system power off. The cooling unit should be activated if used with the system.
4.Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or failure.If the serious problems caused by the above risk is expected,you are requested to have a contingency plan to avoid such a case.
5.New Application
If you use the product with new application not to be mentioned in this specification or with different type of X-ray generator,please contact to us for confirming its availability.
1 .X-ray Radiation protection
This product fulfills the requirements of IEC 60601-1-3.
This X-ray tube assembly emits X-ray radiation in operation.Only correspondingly qualified and trained personnel are therefore allowed to operate the X-ray tube assembly.
Relevant physiologic effects may cause harm to patient,system manufacture should take proper protection to avoid ionization radiation.
2.Dielectric 0il
X-ray tube assembly has dielectric 0il contained for high voltage stability. As it is poisonous for human health,if it is exposed to the non-restricted area,it should be disposed as following to the local regulation.
3 .Operation Atmosphere
X-ray tube assembly is not allowed to be used in the atmosphere of flammable or corrosive gas·
4.Adjust the Tube Current
Depending on the operating conditions,the filament characteristics might be changed.
This change might Iead to the over rate exposure to X-ray tube assembly.
To prevent the X-ray tube assembly from being damaged,adjust the tube current regularly.
Besides when the X-ray tube has arcing Problem in a long time use,adjustment of the tube current is required.
5.X-ray Tube Housing Temperature
Do not touch on X- ray tube housing surface just after operation due to high temperature.
Stay X-ray tube to be cooled.
6 .Operating limits
Prior to usage,please confirm the environmental condition is within the operating Iimits.
7 .Any Malfunction
P1ease contact to SAILRAY immediately,if any malfunction of the X-ray tube assembly is noticed.
8.Disposal
The X-ray tube assembly as well as the tube contain materials such as oil and heavy metals for which environmentally friendly and proper disposal in accordance with the valid national legal regulations must be assured.Disposal as domestic or industrial refuse is forbidden.The manufacturer possesses the required technical knowledge and will takes the X-ray tube assembly back for disposal.
Please contact customer service for this purpose.
Conditions:Tube Voltage Three-Phase
Stator Power Frequency 50Hz/60Hz
IEC60613
Housing Thermal Characteristics
SRMWHX7010A